Specifications

214 Chapter 16
Command Reference
:CALC:COMP:BIN{1-13}:COND{1-4}:LTYP
:CALC:COMP:BIN{1-13}:COND{1-4}:LTYP
Syntax :CALCulate:COMParator:BIN{1|2|3|4|5|6|7|8|9|10|11|12|13}:CONDition{1|2|3|4}:LTYPe {IN|OUT|ALL}
:CALCulate:COMParator:BIN{1|2|3|4|5|6|7|8|9|10|11|12|13}:CONDition{1|2|3|4}:LTYPe?
Description Lets you set the limit range mode of each of the four sorting conditions (:COND1 through
:COND4) for bin 1 (:BIN1) through bin 13 (:BIN13).
Parameters
Query response {IN|OUT|ALL}<newline><^END>
See also :CALC:COMP:BIN{1-13}:COND{1-4}:LIM on page 213
Equivalent key
sequence
[Setup View] (Bin sort setup display)
:CALC:COMP:CLE
Syntax :CALCulate:COMParator:CLEar
Description Initializes (resets) all the sorting conditions for each comparator bin and a limit range for
R
dc
measurement. (No query)
Equivalent key
sequence
[Alt] - [2] - [8] - [2] (when the bin sort setup display is selected by [Setup View])
Description
IN Instructs the instrument to determine that a test has passed the sorting
condition where DUTs fall within the limit range (inclusive of the
upper and lower limits).
OUT Instructs the instrument to determine that a test has passed the sorting
condition where DUTs do not fall within the limit range (exclusive of
the upper and lower limits).
ALL
(initial value)
Instructs the instrument to determine that a test has passed the sorting
condition where the measurement result is disregarded.