Specifications
Chapter 16 213
Command Reference
:CALC:COMP:BIN{1-13}:COND{1-4}:LIM
16. Command Reference
:CALC:COMP:BIN{1-13}:COND{1-4}:LIM
Syntax :CALCulate:COMParator:BIN{1|2|3|4|5|6|7|8|9|10|11|12|13}:CONDition{1|2|3|4}:LIMit <numeric 1>,<numeric 2>
:CALCulate:COMParator:BIN{1|2|3|4|5|6|7|8|9|10|11|12|13}:CONDition{1|2|3|4}:LIMit?
Description Lets you specify the limit range that applies to each of the four sorting conditions
(:COND1 through :COND4) for bin 1 (:BIN1) through bin 13 (:BIN13). The limit range is
inclusive of both upper and lower limit values.
The limit range is applied when the parameter specified with the
:CALC:COMP:COND{1-4}:PAR command is measured at the point specified with the
:CALC:COMP:COND{1-4}:SNUM command.
If you have issued the :CALC:COMP:COND{1-4}:MODE by specifying “DEV” or
“PCNT” to configure the instrument to accept a limit range in deviation or percent mode,
the instrument assumes that the limit values are relative to the reference value specified
with the :CALC:COMP:COND{1-4}:NOM command.
Parameters
If your specified parameter is beyond the valid range, the parameter is replaced with either
the maximum or minimum value of the valid range depending on whether the parameter is
greater than the upper limit or smaller than the lower limit.
Query response {numeric 1}, {numeric2}<newline><^END>
The returned data always lists the lower and upper limits in that order.
See also :CALC:COMP:COND{1-4}:SNUM on page 218
:CALC:COMP:COND{1-4}:PAR on page 217
:CALC:COMP:COND{1-4}:MODE on page 215
:CALC:COMP:COND{1-4}:NOM on page 216
:CALC:COMP:BIN{1-13}:COND{1-4}:LTYP on page 214
Equivalent key
sequence
[Setup View] (Bin sort setup display)
<numeric 1> <numeric 2>
Description
Lower limit value
*1
Upper limit value
*1
Range -1E20 to 1E20 -1E20 to 1E20
Default 0 0
Unit Depends on the sort parameter. Depends on the sort parameter.
Resolution 1E-20 1E-20
*1. If <numeric 1> is greater than <numeric 2>, the instrument uses <numeric 1> as the
upper limit and <numeric 2> as the lower limit.