Specifications
Chapter 16 211
Command Reference
:CALC:COMP
16. Command Reference
:CALC:COMP
Syntax :CALCulate:COMParator[:STATe] {ON|OFF|1|0}
:CALCulate:COMParator[:STATe]?
Description Turns on or off the comparator. This setting is linked with the on/off status of the handler
interface.
Parameters
Query response {1|0}<newline><^END>
Equivalent key
sequence
[System] - COMPARATOR
:CALC:COMP:BEEP:COND
Syntax :CALCulate:COMParator:BEEPer:CONDition {FAIL|PASS}
:CALCulate:COMParator:BEEPer:CONDition?
Description This command is available when the instrument is configured to beep based on the
comparator result (i.e., after you have issued the :SYST:BEEP2:STAT command by
specifying “ON”). The command lets you specify whether the instrument should beep
when the DUT fails to meet the comparator sorting criteria (not sorted into any of bins 1
through 13, or sorted into the NO GOOD bin specified with the :CALC:COMP:OGB
command) or when the DUT satisfies the comparator sorting criteria (sorted into the
GOOD bin).
Parameters
Query response {FAIL|PASS}<newline><^END>
See also :SYST:BEEP2:STAT on page 303
:CALC:COMP:OGB on page 223
Equivalent key
sequence
[System] - BEEPER MENU - BEEP COMPARATOR
Description
ON or 1 Turns on the comparator
OFF or 0 (initial value) Turns off the comparator
Description
FAIL (initial value) Instructs the instrument to beep for a DUT that does not meet the
sorting criteria.
PASS Instructs the instrument to beep for a DUT that meets the sorting
criteria.