Specifications

210 Chapter 16
Command Reference
:AVER:COUN
:AVER:COUN
Syntax [:SENSe]:AVERage:COUNt <numeric>
[:SENSe]:AVERage:COUNt?
Description This command sets the averaging factor.
The measurement point for single point measurement, specified with the
:SOUR:LIST:POIN command, in the active table, specified with the
:SOUR:LIST:TABL command, must be set here.
Execution of this command does not turn off calibration or compensation automatically. If
you need to change only the averaging factor after completion of calibration or
compensation, use this command.
NOTE To change only the averaging factor after completion of calibration/compensation, you
may execute the :SOUR:LIST command by specifying the parameter so that only the
averaging factor will change with the frequency and signal source level identical to the
current settings. In this case, calibration and compensation will be turned off automatically;
you will need to rerun calibration and compensation.
Parameters
If your specified parameter is beyond the valid range, the instrument ignores the parameter
and uses either the maximum or minimum value depending on whether the parameter is
greater than the upper limit or smaller than the lower limit.
Query response {numeric}<newline><^END>
See also :SOUR:LIST on page 294
:SOUR:LIST:TABL on page 297
:SOUR:LIST:POIN on page 295
Equivalent key
sequence
[Setup View] (Measurement Point Setup display)
<numeric>
Description Averaging factor
Range 1 to 100
Initial value 1
Resolution 1