Specifications
192 Chapter 15
Sample Application Programs
Measuring Array (Multi-Element) Components
Measuring Array (Multi-Element) Components
When measuring an array component with the scanner, you must select appropriate
calibration coefficients for each channel, switched by the scanner, to make a correct
measurement (by eliminating errors that depend on measurement paths).
The 4287A can handle up to 8 tables for different measurement conditions (including
calibration coefficients). This capability is used to set calibration coefficients for each
channel. Specifically, calibration coefficients for all of the channels are first obtained in
advance based on tables configured for the same measurement condition. Then, while
performing measurement for each channel, the corresponding table is used. In this way,
you can perform measurements under the same measurement condition but with different
calibration coefficients.
The following describes how to use the 4287A to measure array components.
Step 1. Configuring the setup tables
Configure the measurement point setup tables and the calibration kit setup tables for all
required channels. If you use the bin sorting function (comparator), also configure the bin
sorting setup tables.
Step 2. Obtaining calibration coefficients
Change the channel and active table together and measure the OPEN/SHORT/LOAD data
to calculate the calibration coefficients for each channel.
Figure 15-1 Timing Chart of Array Component Measurement (using 2 channels)