Specifications

108 Chapter 7
Sorting DUTs Based on Measurement Results
Setting Up the Bin Sorting Function
Figure 7-1 Commands for defining sorting conditions
where and mean the following:
: A value obtained by converting the impedance value measured at the
point specified via the :CALC:COMP:COND{1-4}:SNUM
command into the measurement parameter value specified via the
:CALC:COMP:COND{1-4}:PAR command.
: The reference value specified via the
:CALC:COMP:COND{1-4}:NOM command.
Table 7-1 Relationship between limit mode and measured value used for sorting
Limit mode Absolute mode (ABS)
Deviation tolerance
mode (DEV)
Percent tolerance mode
(PCNT)
Measured value
Meas Meas Nom
Meas Nom
Nom
--------------------------------
100´
Meas Nom
Meas
Nom
Table 7-2 Relationship between BIN range mode and test result (PASS/FAIL)
Range of measured value
BIN range mode
In Out All
Measured value < lower limit FAIL PASS
PASSLower limit £ measured value £ upper limit PASS FAIL
Measured value > upper limit FAIL PASS