Specifications

Chapter 6 101
Retrieving Measurement Results
Retrieving the Results of Rdc Measurement
6. Retrieving Measurement
Results
Retrieving the Results of R
dc
Measurement
Retrieving measurement results
To retrieve only the results of R
dc
measurement, use the following command:
:DATA:RDC? on page 267
As described in “Retrieving the Measurement Results for Measurement Parameters 1
through 4” on page 85, you can also use the following three commands to retrieve the Rdc
measurement results along with other data such as the measurement results for
measurement parameters 1 through 4:
*TRG on page 208
:FETC? on page 280
:READ? on page 293
Retrieving limit test results
To determine whether the R
dc
measurement results fall within the limit range (defined with
the :CALC:COMP:RDC:LIM command), use the following command:
:CALC:COMP:DATA:RDC? on page 222
Also, you can use the *TRG or similar command to retrieve the measurement results as
well as the measurement status and then obtain the R
dc
measurement limit test results from
that measurement status.
As shown below, the measurement status is 0 or 1 when the R
dc
measurement results fall
within the limit range and 2 or 3 when the R
dc
measurement results do NOT fall within the
limit range.
Value of Measurement Status
0: Normal
1: Measurement failure (overload)
2: R
dc
measurement result did not fall within the limit range
3: Measurement failed and DUT did not fall within R
dc
limit.