Instruction manual

Chapter 4 85
Function Overview
Names and Functions of Display Areas on the LCD Display
4. Function Overview
you can turn the rotary knob instead of pressing the keys and then
press the rotary knob.
Activate the menu bar by pressing the [Alt] key of the keyboard connected to the
4287A and perform any of the following operations (or combined operations):
Enter the menu name number using the numeric entry keys ([0] to [9]) of the
keyboard to select the desired function.
Move the cursor on the menu using the arrow keys ([], [], [] and []) of the
keyboard and then press the
[Enter] key to select the desired function.
NOTE You can release the state by again pressing the menu bar that was activated by pressing the
[Alt] key on the front panel or keyboard.
For the measurement display menu bar, despite whether one point measurement display or
list measurement display is shown, the function (menu contents) does not change
(common).
3. Table number
Displays the currently selected table number. The 4287A has eight tables that each allow a
maximum of 32 measurement points to be set. Use the
“Measurement point setup display”
on page 91 to set the measurement point conditions (measurement frequency, averaging
factor, oscillator level for the measurement point number) for each table.
NOTE To directly call up the table number selection softkey menu, use the mouse to click in the
table number area.
4. Measurement point number
Displays the currently selected measurement point number. Use the “Measurement point
setup display” on page 91 to set the measurement point conditions (measurement
frequency, averaging factor, oscillator level for the measurement point number) for each
table.
NOTE To directly call up the measurement point number selection softkey menu, use the mouse to
click in the measurement point number area.
5. BIN sort result
When the BIN sort function is turned ON, the sorting result (BIN number) based on the
measurement results is displayed at this position. For the BIN sort function, see
Chapter 8,
“Sorting According to Measurement Results,” on page 171.
6. Measurement frequency
Indicates the measurement frequency of the currently selected measurement point. The
conditions (measurement frequency of measurement point, averaging factor, oscillator
level for the measurement point number) for the measurement points of each table are set
in
“Measurement point setup display” on page 91.