Instruction manual
Chapter 3 69
Learning Basic Operations
Display of DUT Connection and Measurement Result
3. Learning Basic
Operations
Display of DUT Connection and Measurement Result
DUT connection and measurement
Step 1. Connect the DUT according to the Operation Manual for the test fixture.
The measurement result is displayed as shown in Figure 3-28.
Figure 3-28 Measurement result display (measurement point number = 1)
How to change measurement point
In Figure 3-28, the measurement result is only displayed for the 1-MHz point among the
four measurement points (1 MHz, 10 MHz, 100 MHz and 1 GHz) set according to “How to
set the measurement point (frequency/averaging factor/oscillator level)” on page 54.
Follow these steps to change the measurement point on which you want to perform
measurement
s and display results.
Step 1. Press the [Stml Select] key on the measurement block (MEASUREMENT) of the front
panel. The Stml Select softkey menu appear
s on the right side of the display.
Step 2. Turn the rotary knob ( ) in the navigation block (NAVIGATION) on the front panel to
move the cursor in the softkey menu to the POINT No.[ ] key.
Step 3. Press the rotary knob ( ) to display the POINT No.[ ] softkey menu on the right side of
the display.
Step 4. Turn the rotary knob ( ) to move the cursor in the softkey menu to the softkey
corresponding to the desired measurement point number (POINT 4 in this case).