Instruction manual
Chapter 3 65
Learning Basic Operations
Calibration and Compensation
3. Learning Basic
Operations
Connection of Test Fixture
Connect the test fixture as follows:
Figure 3-22 Test fixture connection window
Step 1. Turn the connector nut of the low-loss capacitor counterclockwise and remove it from the
7-mm terminal.
Step 2. Turn the connector nut of the 7-mm connector counterclockwise to retract the connector
sleeve completely inside
(Figure 3-23).
Figure 3-23 Retracting the 7-mm connector connector sleeve
Step 3. Fit the tips of the two mounting posts on the test fixture stand to the two holes of the test
fixture
and fit the 7-mm connector on the test fixture stand to the 7-mm connector on the
test fixture
. Turn the connector nut of the 7-mm connector on the test fixture stand
counterclockwise
, as viewed from above, and firmly lock it to the test fixture’s 7-mm
connector.