64 Chapter 3
Learning Basic Operations
Calibration and Compensation
Figure 3-20 How to retract connector sleeve of load standard
Step 3. While pressing the load standard (50 Ω) from above, turn the connector nut clockwise and
s
ecurely connect it to the 7-mm terminal (Figure 3-21).
Figure 3-21 How to connect load standard
Step 4. Press the key or key on the front panel the required number of times to move the
cursor to the Next> button.
Step 5. Press the key on the front panel to start measurement of load calibration data. The bar
in the Calibration progress indicator in the calibration window indicates the progress of the
load calibration data measurement. Upon completion of load calibration data measurement,
the screen changes to the next window (
Figure 3-22).