Instruction manual

Chapter 3 63
Learning Basic Operations
Calibration and Compensation
3. Learning Basic
Operations
Figure 3-18 How to connect low-loss capacitor
Step 3. Press the key or key on the front panel the required number of times to move the
cursor to the Next> button.
Step 4. Press the key on the front panel to start measurement of low-loss capacitor calibration
data. The bar in the Calibration progress indicator in the calibration window indicates the
progress of the low-loss capacitor calibration data measurement. Upon completion of
low-loss capacitor calibration data measurement, the screen changes to the next window
(
Figure 3-19).
Executing load calibration
Figure 3-19 Load calibration window
Step 1. Turn the connector nut of the low-loss capacitor counterclockwise to remove it from the
7-mm terminal.
Step 2. As shown in Figure 3-20, turn the outside connector nut of the load standard (50 Ω)
counterclockwise to fully retract the internal connector sleeve.