Chapter 3 61
Learning Basic Operations
Calibration and Compensation
3. Learning Basic
Operations
Figure 3-14 How to connect short standard
NOTE After turning the connector nut of the short standard and attaching it to the 7-mm terminal,
it is not necessary to turn the connector nut of the 7-mm terminal on the test head side to fix
it in position. This also applies to the installation
s of the open standard, low-loss capacitor
and load standard explained in this section.
Step 11. Press the key or key on the front panel the required number of times to move the
cursor to the Next> button.
Step 12. Press the key on the front panel to start measurement of short calibration data. The bar
in the Calibration progress indicator in the calibration window indicates the progress of the
short calibration data measurement. Upon completion of short calibration data
measurement, the screen changes to the next window (
Figure 3-15).
Executing open calibration
Figure 3-15 Open calibration window
Step 1. Turn the outside of the short standard counterclockwise and remove it from the 7-mm
terminal.