60 Chapter 3
Learning Basic Operations
Calibration and Compensation
Step 7. Press the key or key on the front panel the required number of times to move the
cursor to the Next> button.
Step 8. Press the key on the front panel. The window for short calibration (Figure 3-12)
appears.
Figure 3-12 Window for short calibration
Step 9. The outside of the 7-mm connector of the test head installed on the test fixture stand should
be turned clockwise until the connector sleeve protrudes
to its full extent at the top (Figure
3-13).
Figure 3-13 How to extend 7-mm connector sleeve
Step 10. While pressing on the short standard (0 Ω) from the top, turn it clockwise and securely
connect it to the 7-mm terminal (
Figure 3-14).