Instruction manual

Chapter 3 57
Learning Basic Operations
Calibration and Compensation
3. Learning Basic
Operations
Calibration and Compensation
After completing measurement point setting, perform calibration and compensation
NOTE Calibration and compensation are carried out at the same point as the normally set
measurement point; therefore, the measurement point setting should have been completed
before calibration and compensation.
The following description uses the wizard function, which ensures easy execution of
calibration and compensation
by displaying clear and simple guidance.
Low-loss capacitor calibration, port extension compensation, open compensation, and
short compensation in addition to open/short/load calibration
are explained in the
measurement
examples below.
The definition of the calibration kit and compensation kit remains the same as that in the
presetting phase.
Setting with the calibration and compensation wizard
Perform the following steps before using the calibration and compensation wizard to make
settings.
NOTE When using test fixtures other than 16191A, 16192A, 16193A, 16194A, 16196A, 16196B,
or 16196C, register the electrical length as described in
“Registering the Electrical Length
of Unregistered Test Fixtures” on page 149 before making settings using the calibration
and compensation wizard.
Step 1. Press the [Cal/Compen] key in the instrument state block (INSTR STATE) on the front
panel to display the Cal/Compen softkey menu in the right side of the screen.
Step 2. Turn the rotary knob ( ) in the navigation block (NAVIGATION) on the front panel to
move the cursor in the softkey menu to the SETUP CAL WIZARD key.
Step 3. Press the rotary knob ( ) to display the Calibration Wizard Setting dialog box (Figure
3-9).
Figure 3-9 Calibration Wizard Setting dialog box