Instruction manual

326 Appendix F
Error Messages
Save error
S
104 Save error
When saving a file, anomalies in the storage media were detected. For example, if you
attempt to save a file on a floppy disk when there is not enough space on the disk.
Ensure that there is enough space on the floppy disk.
-221 Setting conflict
A program data element complying with the syntax standard was analyzed but the 4287A
could not execute it at the time.
-150 String data error
An error not included in error numbers between -151 and -159 occurred during the syntax
analysis of a string data element.
-158 String data not allowed
A string data element was received where the 4287A does not accept any string data
element. For example, a parameter that should not be enclosed with double quotation
marks (“) such as ABC is sent with them, as in “ABC”.
-138 Suffix not allowed
A suffix is added to a numeric value element to which no suffix can be added.
-134 Suffix too long
The suffix contains the notation of 12 characters or more. (Refer to IEEE488.2,7.7.3.4.)
-102 Syntax error
There is a command or data type that cannot be recognized. For example, for a correct
program message, “:SYST:PRES”, an extra colon (:) is inserted by mistake, as in
:SYST::PRES”.
-310 System error
One of the “system errors” defined for the 4287A occurred.
T
-124 Too many digits
The number of digits of the mantissa of the decimal value data element exceeds 255, aside
from preceding 0s. (Refer to IEEE488.27.7.2.4.1.)
-223 Too much data
The received block, equation, or string type program data complied with the standard but
its amount exceeded the limit that the 4287A can deal with due to memory or
device-specific conditions related to memory.
-211 Trigger ignored
A trigger command or trigger signal was received and recognized by the 4287A but was
ignored due to the timing relationship with the 4287A (for example, when the 4287A’s
trigger system is not in Waiting for Trigger state).