Instruction manual

Appendix F 325
Error Messages
Power on test failed
1. Chapter Title
2. Chapter Title F. Error Messages 5. Chapter Title
198 Power on test failed
In the power-on self test, a fault was detected.
Contact your local Agilent Technologies sales office or the company you purchased this
instrument from.
22 Printer error
The printer does not respond to control from the 4287A.
Check the power to the printer, cable connections, paper, and so on.
-112 Program mnemonic too long
The length of the header exceeds 12 characters. (Refer to IEEE488.2,7.6.1.4.1.)
Q
-430 Query DEADLOCKED
Status that causes a “DEADLOCKED” query error. (Refer to IEEE488.2,6.3.1.7.) This
error occurs, for example, when both input and output buffers become full and the 4287A
cannot continue processing.
-400 Query error
A comprehensive Query error occurred for which the 4287A could not detect further
details. This code simply indicates the occurrence of a query error that is defined in
IEEE488.2,11.5.1.1.7 and 6.3.
-410 Query INTERRPUTED
Status that causes an “INTERRUPTED” query error. (Refer to IEEE488.1,6.3.2.3.) This
error occurs, for example, when data byte (DAB) or GET is received after Query has been
initiated but before the response has been completely sent.
-420 Query UNTERMINATED
Status that causes an “UNTERMINATED” query error. (Refer to IEEE488.2,6.3.2.) This
error occurs, for example, when the 4287A is specified as a talker and an incomplete
program message is received.
-440 Query UNTERMINATED after indefinite response
In a certain program message, after a Query that requested an ambiguous response was
executed, another Query was received. (Refer to IEEE488.2,6.5.7.5.7.)
R
105 Recall error
An error occurred while reading out (recalling) a file. For example, you attempted to read
out a file with invalid contents, such as an instrument setting file with extension “.sta” that
was saved using an instrument other than the 4287A).
Check to see if the contents of the file is valid.