Instruction manual

312 Appendix D
Table of Operations When Measurement Errors Occur
Operations When Measurement Errors Occur
Operations When Measurement Errors Occur
Table D-1 shows the LCD screen indications, GPIB output, and the parameters and
interface signals that become active when any of the following items is detected.
Overload
(Detected for ranging failure caused by DUT coming loose during measurement, etc.)
•R
dc
limit deviation
(Detected when the R
dc
measurement result does not fall within the limit range.)
Simultaneous detection of overload and R
dc
limit deviation.
Table D-1 Operations when errors occur
LCD screen indication GPIB output
Parameter and interface
signal that become active
Measurement result
Comparator
sorting
result
Measurement
status
Measurement
value
*1
Comparator
sorting
result
Parameter
1 to 4
Measurement
signal level
monitor
R
dc
Overload
*2
9.9E37 9.9E37
Measurement
value
*3
14
1 9.9E37
14
/OVLD
R
dc
limit deviation
Measurement
value
*4
Measurement
value
*4
Measurement
value
*5
(indicated by
red
characters)
2
Measurement
value
*6
/RDC_OUT_OF_RANGE
Simultaneous
detection of
Overload
*2
and
R
dc
limit deviation
9.9E37 9.9E37
Measurement
value
*5
(indicated by
red
characters)
3 9.9E37
/OVLD
/RDC_OUT_OF_RANGE
*1. All measurement values of parameters 1 to 4 (impedance measurement), measurement signal level monitor, and
R
dc
measurement.
*2. Ovld lights up red in the instrument status area at the bottom of the display.
*3. Overload is not detected for R
dc
measurement.
*4. The measurement result is indicated as usual, but the indicated value may have been measured in a condition of
improper contact.
*5. The measurement result is indicated as usual, but the characters are displayed in red.
*6. The measurement result is read out as usual, but the readout value may have been measured in a condition of
improper contact.