Instruction manual

304 Appendix B
Differences from Agilent 4286A
3. Functional Differences Between 4286A and 4287A
Sweep Setup SEGMENT Set Segment
EDIT
DELETE
ADD Set Frequency within SEGMENT
FREQ Segment
Set AVG within AVERAGING ON
POINT
SEGMENT QUIT
SEGMENT DONE
CLEAR LIST Clear YES/NO List Sweep Tables
For the 4287A, tables can be cleared one by
one, or all tables can be cleared at once.
LIST DONE Complete Setting of List Sweep
Tables
Test Setup LIMIT TEST
MENU
Limit Test of PRI LIMIT ON/OFF
Primary Parameter
The 4287A is not provided with the limit test
function as provided for the 4286A.
Limit Test of PRI LIMIT ON/OFF
Secondary Parameter
Input Limit of EDIT PRI LIMIT
Primary Parameter
Input Limit of EDIT SEC LIMIT
Secondary Parameter
BIN SORT
MENU
BIN SORT ON/OFF Comparator
ON/OFF
The concept of the comparator function for
the 4287A is completely different from that
for the 4286A.
Select TEST SEGMENT
Limit Setting of EDIT BIN SORT
Bin Sort
BEEP Select the setting to generate a beep
when a test result is PASS.
Select the setting to generate a beep
when a test result is FAIL.
HANDLER IF Output
MODE:
The 4286A is provided with two output lines
for MODE1 and MODE2; the 4287A is
provided with one output mode because it is
based on a comparator concept that is
completely different from that of the 4286A.
ON/OFF:
The 4286A can be set up independently.
The 4287A is dependent on the ON/OFF of
the comparator function and cannot be set up
independently.
Monitor AC-V AC-V Monitor Display
AC-I AC-I Monitor Display
Table B-5 Measurement functions
Function 4286A 4287A Remarks