Instruction manual

302 Appendix B
Differences from Agilent 4286A
3. Functional Differences Between 4286A and 4287A
Calibration CALIBRATION
MENU
Run OPEN The concept of calibration/compensation for
the 4286A differs from that for the 4287A.
In addition, for the 4286A, the compensation
data for Rdc measurement is obtained
together with that for calibration; for the
4287A, these kinds of data can be obtained
separately.
Run SHORT
Run LOAD
Run LOW-LOSS Capacitor
Select CAL POINT [FIXED]
Interpolation Mode
Because the interpolation mode is not
supported for the 4287A, the FIXED mode
is not available; it is always in user mode.
Select CAL POINT [USER] User
Mode
RESUME CAL
SEQUENCE
For the 4287A, even when the Run screen
closes temporarily in the course of
calibration/compensation, the
calibration/compensation data already
processed is not cleared.
Calibration FIXTURE
COMPEN
Run COMPEN OPEN
Run COMPEN SHORT
Run COMPEN LOAD
For the 4287A, LOAD compensation is not
available; calibration should be used for this
purpose.
Select COMPEN POINT [FIXED]
Interpolation Mode
Because the interpolation mode is not
supported for the 4287A, the FIXED mode
is not available.
Select COMPEN POINT [USER]
User Mode
The 4287A is always in user mode.
RESUME COM SEQUENCE Data
Resume
For the 4287A, even when the Run screen
closes temporarily in the course of
calibration/compensation, the
calibration/compensation data already
processed is not cleared.
OPEN Compensation ON/OFF
SHORT Compensation ON/OFF
LOAD Compensation ON/OFF
For the 4287A, LOAD compensation is not
available; calibration should be used for this
purpose.
FIXTURE Select SELECT FIXTURE 16191
The 4287A is additionally provided with
16196A/B/C.
Select SELECT FIXTURE 16192
Select SELECT FIXTURE 16193
Select SELECT FIXTURE 16194
Select SELECT FIXTURE 16196A
Select SELECT FIXTURE 16196B
Select SELECT FIXTURE 16196C
Table B-5 Measurement functions
Function 4286A 4287A Remarks