Instruction manual
Chapter 12 271
Specifications and Supplemental Performance Characteristics
Measurement Support Functions
12. Specifications and
Supplemental Performance
Characteristics
Trigger function
Averaging function
Display
List measurement function
Test signal level monitor function
Mass Storage
Interface
GPIB
Trigger mode Internal, External (external trigger input connector or
handler interface), Bus (GPIB or LAN), Manual (front key)
Setting range 1 to 100 (integer)
Type/Size Color LCD, 8.4 inch
Resolution 640 dots × 480 lines
Number of Measurement points 32 points for each table (maximum)
Number of tables 8 tables
Uncertainty of monitor value
[%](SPC)
A: Uncertainty of oscillator level [dB]
B: Uncertainty of impedance measurement [%]
Built-in flexible disk drive 3.5 inch, 720 KByte or 1.44 KByte, DOS format
Built-in hard disk drive About 18 GByte
Standard conformity IEEE 488.1-1987, IEEE 488.2-1987
Available functions
(function code)
*1
*1. Refer to the standard for the meaning of each function code.
SH1, AH1, T6, TE0, L4, LE0, SR1, RL0, PP0, DT1, DC1,
C0, E2
Numerical data transfer format ASCII
Protocol IEEE 488.2-1987
30 10
A
20
------
1–
100× B++±