Instruction manual
Chapter 12 269
Specifications and Supplemental Performance Characteristics
Basic Measurement Characteristics
12. Specifications and
Supplemental Performance
Characteristics
Measurement Data Transfer Time through GPIB (Typical)
Conditions:
4287A: S/N JP1KGxxxxx
Host computer: HP9000 Series, Model 715
Display: OFF
Measurement mode: List measurement
Measurement parameters: Ls and Q
Evaluation method: Average value of 10,000 repeated measurements
Switching Time for Setup Change by GPIB (Typical)
Conditions:
Host computer: HP9000 Series, Model 715
Display: OFF
Measurement mode: List measurement
Measurement parameters: Ls and Q
Evaluation method: Average value of 10,000 repeated measurements
Table 12-4 Measurement Data Transfer Time through GPIB (Typical)
Number of measurement
points
Required time for FETCH? command
Rdc measurement OFF Rdc measurement ON
1 10.0 ms 10.0 ms
2 12.0 ms 13.5 ms
3 15.0 ms 15.0 ms
Table 12-5 Switching Time for Setup Change by GPIB (Typical)
Conditions Time
Measurement table switching at list
measurement (required time for executing
:SOUR:LIST:TABL 1 command and
*OPC? command)
8.8 ms