Instruction manual

192 Chapter 8
Sorting According to Measurement Results
Example of Settings for BIN Sorting
Example of Settings for BIN Sorting
The following introduces an example of setting an actual BIN sort setting table. The
sorting conditions to be set are shown in
Table 8-2.
NOTE When both Condition 1 and Condition 2 of Table 8-2 are met, the DUT is placed in the
appropriate BIN.
In Table 8-2, “” denotes that this point is included, and “” denotes that this point is not
included.
Perform the following steps to set these conditions in the 4287A.
Table 8-2 Sorting conditions
BIN Condition 1 Condition 2
Measurement point number: 3 (100 MHz)
Measurement parameter: Ls
Measurement point number: 3 (100 MHz)
Measurement parameter: Q
BIN 1
BIN 2
BIN 3
OUT OF
BINS