Instruction manual
Chapter 8 191
Sorting According to Measurement Results
Bin Count Function
8. Sorting According to
Measurement Results
Figure 8-16 Bin Count Results dialog box
NOTE To close the Bin Count Results dialog box, press the OK button.
The BIN Count results listed in the dialog box are those obtained when you pressed the
DISPLAY softkey. If the contents of the bin count are updated because measurement and
bin sorting are run when the dialog box is on display, you will have to close it once and
then press the DISPLAY softkey again to redisplay it.
Initializing the contents of the bin counts to zero (resetting the counter)
The contents of all bin counts (i.e., value in the counter) can be initialized (resetting) to
zero using the following procedure:
Step 1. In the front panel Measurement Block, press the [Prmtr] key to display the Prmtrs softkey
menu.
Step 2. In the softkey menu, press the BIN COUNT MENU to display the BIN COUNT MENU
softkey menu.
Step 3. In the softkey menu, press the CLEAR key to initialize the contents of all bin counts to
zero.