Instruction manual
Chapter 8 185
Sorting According to Measurement Results
Setting BIN Range
8. Sorting According to
Measurement Results
Selecting BIN range mode
You can select whether the BIN range should be inside or outside of the set limit values
(upper and lower limit values) or whether the BIN range should cover the entire range
regardless of the limit values.
Step 1. Click with the mouse in the cell corresponding to the column of the desired sorting
condition (one of conditions 1 to 4) on the In/Out row of the BIN to be set.
Step 2. Click the down arrow () at the right edge of the cell.
Step 3. Click the desired BIN range mode from the drop-down list box to select it.
Figure 8-11 Selecting BIN range mode
Bin range mode Definition
In Inside. This mode ensures that the BIN
range is inside the set limit values (upper
and lower limit values).
Out Outside. This mode ensures that the BIN
range is outside the set limit values (upper
and lower limit values). Namely, the BIN
is used when the result is equal to or
greater than the upper limit value or equal
to or less than the lower limit value.
All
The entire range. In this mode the BIN
range covers the entire range regardless of
the set limit values (upper and lower limit
values). Namely, the system assumes that
any measurement result is within the set
range.