Instruction manual

Chapter 8 183
Sorting According to Measurement Results
Setting BIN Range
8. Sorting According to
Measurement Results
Setting BIN Range
Set the actual BIN range after setting the BIN range setting conditions according to
“Setting Measurement Point” on page 177), “Setting Measurement Parameter” on page
179, and “Setting Limit Mode and Reference Value” on page 181.
Setting upper limit value
Follow these steps to set the BIN upper limit value.
Step 1. Click with the mouse the OFF button under the BIN to be set. This turns on the BIN and
changes the button to
ON.
Step 2. Click in the cell corresponding to the column of the desired sorting condition (one of
conditions 1 to 4) on the row of the upper limit value of the BIN to be set.
Step 3. Press the entry block numeric keys ([0] to [9][-][.]) to enter the upper limit value of the BIN
to be set and then press the desired unit key (
[G/n], [M/μ], [k/m], or [×1]).
To set the upper limit value to 51 nH when the limit mode is in absolute mode, for example,
press
[5], [1], and then [G/n].
To set the upper limit value to +1% of the reference value (Nominal) when the limit mode
is in percentage mode, press
[1] and then [×1].(Figure 8-9)
Figure 8-9 Setting upper limit value of BIN 1