182 Chapter 8
Sorting According to Measurement Results
Setting Limit Mode and Reference Value
Setting reference value
When you select percentage mode (PCNT) or deviation mode (DEV) as the limit mode, you
must set the reference value (Nominal).
NOTE When you select absolute mode (ABS) as the BIN limit mode, you do not need to set the
reference value (nominal). Actually, this value cannot be set in ABS mode.
Step 1. Click with the mouse in the reference value (Nominal) cell in the sorting condition to be set
(one of conditions 1 to 4). (Before doing this, make sure the limit mode is set as PCNT or
DEV.)
Step 2. Press the entry block numeric keys ([0] to [9][-][.]) to enter the reference value and then
press the desired unit key (
[G/n], [M/μ], [k/m], or [×1]).
To enter 50 nH, for example, press [5], [0], and then [G/n] (Figure 8-8).
Figure 8-8 Entering BIN limit setting reference value for sorting condition 1