Instruction manual

Chapter 8 181
Sorting According to Measurement Results
Setting Limit Mode and Reference Value
8. Sorting According to
Measurement Results
Setting Limit Mode and Reference Value
Setting limit mode
This procedure determines the format used to set the upper and lower limits of the
measurement values for BIN sorting.
Step 1. Click with the mouse in the BIN limit setting mode cell in the sorting condition to be set
(one of conditions 1 to 4).
Step 2. Click the down arrow () at the right edge of the cell.
Step 3. Click the desired limit mode from the drop-down list box to select it (Figure 8-3).
Figure 8-7 Selection of limit mode for sorting condition 1
Limit mode Definition
ABS Absolute mode. The limits of each BIN
are set as absolute values. If you select
this mode, you do not need to set the
reference value (nominal).
PCNT Percentage mode. The limit value of each
BIN is set as a percentage of the reference
value.
DEV Deviation mode. The limit value of each
BIN is set as a difference with respect to
the reference value. Use the same unit as
that of the set measurement parameter.