Instruction manual
Chapter 7 169
Executing Measurements and Displaying Results
Turning ON/OFF Update the LCD Display View
7. Executing Measurements
and Displaying Results
screen regardless of which command is executed.
2. The display time is much shorter for the list measurement display than for the single
point measurement display
3. The display time becomes shorter as the number of display parameters decreases.
4. If possible, set the screen display to not perform updates (UPDATE:OFF). This
minimizes the time needed to update the screen. For more details, refer to Table 12-4.
Table 7-3 Required measurement conditions and recommended settings
Required measurement
conditions
Recommended settings for shortest
measurement time
Required
measurement
display
parameters
GPIB
command(s)
Screen
display
(DISP)
Display mode
Screen
display
update
(UPDATE)
None Send or None OFF Not relevant Not relevant
All Send ON List
measurement
ON
All None ON List
measurement
ON
Measurement
results only
Send ON List
measurement
OFF
Measurement
results only
None ON List
measurement
OFF