7. Executing Measurements
and Displaying Results
Selection of Measurement Point
Select the measurement point in the single point measurement display as follows:
Step 1. Press the [Stim Select] key in the measurement block to display the Stml Select softkey
menu.
Step 2. Select the POINT No. [ ] key from the softkey menu to display the POINT No.[ ] softkey
menu.
Step 3. Select the measurement point number from the softkey menu.
NOTE Only the measurement points set in the selected measurement point table are listed in the
POINT No.[ ] softkey menu.
Instead of performing the key operation [Stim Select] - POINT No. [ ] as described above,
you can use the mouse to click the measurement point number area (
“4. Measurement point
number” on page 85) in the upper-left corner of the screen to display the POINT No.[ ]
softkey menu.
The list measurement display allows you to select the measurement point in the same
manner. In this case, the selected measurement point number in
“3. Measurement point
number/measurement frequency” on page 93 is highlighted, and the following values for
the selected measurement point are shown at the top of the display: “1. Measurement point
number” on page 89, “6. Measurement frequency” on page 85, “7. Oscillator level
(power)” on page 86, “8. Averaging factor” on page 86, “9. Test signal current level
monitor value” on page 86, and “10. Test signal voltage level monitor value” on page 86.