Instruction manual

148 Chapter 6
Calibration and Compensation
How to Define Compensation Kit
Figure 6-20 Compensation kit setup window
Step 2. Specify the table number to define a compensation kit by following the description in
“How to specify table number for setup” on page 104.
Step 3. Select the type of compensation kit (Default or User Defined) from the Compen Kit
selection area in the compensation kit setup window.
If you have selected “Default,” the initial value of the predetermined compensation kit will
be used as the defined compensation kit. No further setting is required.
If you have selected “User Defined,” continue with the following steps:
Step 4. If you want to use the same definition for all measurement points in the measurement point
table, click the
Fixed Model button to attach a check mark ().
If you have selected “Fixed Model,” set the compensation kit definition for measurement
point number 1. For other measurement points, the same definition as that of measurement
point number 1 is automatically entered.
Step 5. Enter open/short definitions for each measurement point.
Step 6. Enter the DC open/short definitions.