Instruction manual

146 Chapter 6
Calibration and Compensation
How to Define Calibration Kit
Figure 6-18 Calibration kit setup window
Step 2. Specify the table number to define a calibration kit by following the description in “How to
specify table number for setup” on page 104.
Step 3. Select the type of calibration kit (7-mm Standard or User Defined) from the Cal Kit
selection area in the calibration kit setup display.
If you have selected “7-mm Standard,” the initial value of the 7-mm calibration kit will be
used as the defined calibration kit. No further setting is required.
NOTE When you want to used the Agilent 16195B 7-mm calibration kit, select “7-mm
Standard.”
If you have selected “User Defined,” continue with the following steps:
Step 4. If you want to use the same definition for all measurement points in the measurement point
table, click the
Fixed Model button to attach a check mark ().
If you have selected “Fixed Model,” set the calibration kit definition for measurement
point number 1. For other measurement points, the same definition as that of measurement
point number 1 is automatically entered.
Step 5. Select the circuit model to define load from Rs-Ls, Ls-Q and Cp-D.
Step 6. Enter open/short/load definitions for each measurement point.
Step 7. Enter the DC open/short/load definitions.
Step 8. When you want to use a calibration kit other than the 7-mm calibration kit, you may have
to set the offset delay. If so, enter each value for offset delay open/short/load.
NOTE The offset delay value is described in the Operation Manual of the calibration kit. Please
refer to this manual.