Chapter 6 145
Calibration and Compensation
How to Define Calibration Kit
6. Calibration and
Compensation
How to Define Calibration Kit
Before starting calibration, the characteristic values of the open/short/load standards used
for calibration must be set in advance.
For the Agilent 4287A, define the calibration kit by using the circuit model shown in
Figure 6-17.
Figure 6-17 Calibration kit circuit model
Calibration kit definition procedure
Follow these steps to define the calibration kit.
Step 1. Press the [Setup View] key until the calibration kit setup display (Figure 6-18) appears.