124 Chapter 6
Calibration and Compensation
Calibration and Compensation Procedures Using Dedicated Test Fixture
Figure 6-6 How to extend the 7-mm connector sleeve
Step 2. Firmly connect the open standard (0 S) to the 7-mm terminal on the test head by turning it
clockwise as viewed from the top (
Figure 6-7).
Figure 6-7 How to connect open standard
Step 3. Click the OPEN button in the calibration window and measure the open calibration data.
Step 4. Turn the open standard counterclockwise as viewed from the top and remove it from the
7-mm terminal.
6. How to measure short calibration data
Step 1. Firmly connect the short standard (0 Ω) to the 7-mm terminal on the test head by turning it
clockwise as viewed from the top (
Figure 6-8).