Chapter 6 123
Calibration and Compensation
Calibration and Compensation Procedures Using Dedicated Test Fixture
6. Calibration and
Compensation
Figure 6-5 Displaying measured DC resistance value (at calibration data measurement)
To erase all calibration data after measurement is finished, press the Clear Calibration
Data button.
To close the dialog box during calibration, press the Close button.
After turning all of the connector nuts on the standard to secure them to the 7-mm
terminals, it is not necessary to turn the connector nut of the 7-mm terminal of the test
head to secure it.
4. How to select the table range for executing the calibration
Select the table range for executing the calibration by making the appropriate setting in the
Select Calibration Table area of the Calibration Dialog box window.
If you select Active table only, calibration will be executed only for the measurement
point table (active table) selected according to “How to select the measurement point table”
on page
115.
When you select Calibrate All Tables, calibration will be executed for all measurement
point tables.
5. How to measure open calibration data
Step 1. With the test head mounted on the test fixture stand, turn the outer connector nut of the
7-mm connector clockwise as viewed from the top (
Figure 6-6) until the connector sleeve
is fully extended.