Instruction manual
122 Chapter 6
Calibration and Compensation
Calibration and Compensation Procedures Using Dedicated Test Fixture
Step 2. Press the TABLE No. [ ] key in the softkey menu to display the TABLE No. [ ] softkey
menu.
Step 3. In the softkey menu, press the key (TABLE 1 to TABLE 8) corresponding to the
measurement point table for which calibration and compensation data measurement should
be performed.
3. How to display the Calibration Dialog box
Step 1. Press the [Cal/Compen] key on the front panel to display the Cal/Compen softkey menu.
Step 2. Press the CAL DIALOG key in the softkey menu to display the Calibration Dialog box
(
Figure 6-4).
Figure 6-4 Calibration Dialog box
NOTE The following points require attention when performing calibration data measurement.
• When you start the calibration data measurement, a bar appears in the Calibration
progress indicator box of the dialog box to show how the measurement is progressing.
Upon completion of data measurement, a blue mark () appears to the right of the
executed calibration button.
• When the OPEN, SHORT, or LOAD button or the DC OPEN, DC SHORT, or DC
LOAD button is pressed, the value of the DC resistance (Rdc) measured as the
calibration data is shown in the lower-right area of the dialog box upon completion of
the respective measurement (
Figure 6-5). Checking this indication at the time of
calibration data measurement makes it possible to determine that the standard is
properly attached.
• Maximum displaying value of the DC resistance is 100 kΩ.