Specifications

TEST SIGNAL LEVEL ACCURACY TEST
Test
Test Signal
Minimum
Measured
Maximum Measurement
Frequency Level
Value Uncertainty
100
kHz
1 Vrms
0.919 V
1.081
V 0.042 V
1
MHz
5
mVrms
3.580 mV 6.420 mV
0.210
mV
1
MHz
20 mVrms
17.32
mV 22.68 mV 0.84 mV
1
MHz
100
mVrms 90.6
mV
109.4 mV
4.2
mV
1
MHz
1 Vrms
0.915 V 1.085 V 0.042 V
1
MHz
2 Vrms
1.831
V 2.169 V 0.084 V
10
MHz
1
Vrms
0.88 V 1.12 V 0.06 V
30
MHz
1 Vrms
0.80 V 1.20 V 0.10 V
30
MHz
2 Vrms 1.60 V 2.40 V
0.20 V
IMPEDANCE MEASUREMENT ACCURACY TEST
1
pF
STANDARD (0 m)
c.V. :
_____
(Cal. Value at 1 MHz)
Test
Frequency:
1 MHz
Upper
Value: Capacitance (Cp)
Lower
Value:
Dissipation Factor
(D)
Test Signal
Minimum Measured Maximum
Measurement
Level
Value
Uncertainty
5
mV
C.V. -0.0399
pF
C.V.
+0.0399 pF 0.00025 pF
-0.0399 0.0399 0.0002
20
mV
C.V. -0.0113
pF
C.V. +0.0113
pF
0.00025 pF
-0.0113
0.0113 0.0002
100
mV
C.V. -0.0037
pF
C.V. +0.0037 pF 0.00025 pF
-0.0037 0.0037 0.0002
1 V
C.V. -0.0027
pF
C.V. +0.0027
pF
0.00025 pF
-0.0027
0.0027 0.0002
2 V C.V. -0.0022
pF
C.V. +0.0022
pF
0.00025 pF
-0.0022
0.0022
0.0002