Specifications

PERFORMANCE
TESTS
1-16. HANDLER INTERFACE FUNCTION TEST (OPTION 202 ONLY)
Perform this
test
only when troubleshooting the Option 202 handler interface board.
This test verifies the
Option 202 handler interface functions. When this test
is
performed the
following LEOs WILL
NOT
turn
ON
because the signals they represent are not used by the
Option 202 handler interface board.
PHI-, PLO·, SREJ·, UNBAL- and ALARM·
EQUIPMENT:
Handler Simulator
Cable
PROCEDURE:
HP
PN
04278-65001
HP
PN
04278-61635
1. Perform steps 1 through 5 described on page 1-34.
2. Disconnect the
flat cable from the handler interface board. The handler inter-
face board has
ORANGE and
BROWN
extractors and its location is shown in
Figure
1-20 (page 1-34).
3. Remove the
handler interface board.
NOTE
Before performing step
4,
make a note of the jumper settings so that you
can return them to there original settings at the end
of
this test.
4.
Set the jumpers on the handler interface board
as
shown in Figure 1-24.
(Figure 1-24 shows the
jumper settings when the board
is
shipped from the
factory.)
1·38