1-13. VOLTAGE RATIO MONITOR ACCURACY TEST (OPTION 002 ONLY)
This test verifies that the HP 4285A voltage Ratio Monitor accuracy
is
within ± 1 %. The
accuracy is specified for an ambient temperature range
of
23 0 C ±5 0
C.
NOTE
The detailed voltage ratio monitor specification
is
described in the
HP
42851A
Operation
Manual.
HP 4285A Accessory Control
IF
Cable
I
DI
I~
oo~oo
fI
8 M
000
lkJ
o 0 0
CJCJ
CJ
o 0 CJCJoo 0
o
o~
0 0 I = I
BNC(m)-BNC(m)
Cable.
30
cm
BNC(f}-N(m)
.0.
_____
Adapter
BNCIf}-N(m)
~
Adapter
~
A
8-
-u
20
dB Attenuator
d=J
Step Attenuator
HP 42851A
D
Figure 1-16. Voltage Ratio Monitor Accuracy Test Setup
EQUIPMENT:
Q Adapter
Step Attenuator
20 dB Attenuator
BNC(m)-BNC(m) Cable, 30 cm
BNC(f)-N(m) Adapter
PROCEDURE:
HP
42851A
HP
8495A Option 001, H04
HP
8491A Option 020
HP
PN
8120-1838, 2
ea.
HP
PN
1250-0780
1. Set up the equipment
as
shown in Figure 1-16. The H P 4285A is connected to
the
HP
42851A, using the Q Adapter Control Interface Cable.
2. Press the CATALOG/SYSTEM MENU key and the 'SYSTEM CONFIG' softkey to
display the SYSTEM
CON
FIG
page.
1-29