Specifications
PERFORMANCE TESTS
1-12. DC BIAS LEVEL ACCURACY TEST (OPTION
001
ONLY)
This test verifies that the HP 4285A Internal
DC
Bias Level accuracy meets the following
specifications. The accuracy is specified for
an
ambient temperature range of 23 ° C ± 5
°C.
Voltage Range Setting Accuracy
±(O.OOO
to
4.000) V
±(0.1 %
of
setting+1 mY)
±(4.002
to
8.000) V
±(0.1 %
of
setting+2 mY)
±(8.005 to 20.000) V ±(0.1 %
of
setting+5 mY)
±(20.01 to 40.00) V ±(0.1 %
of
setting+ 10 mY)
HP
4285A
MULTIMETER
BNC(f}-dual banana
plug Adapter
liD
I~
ClCl~OCl
ffl
0°00000
CJOOCJCJD
o Q 0
c:::t
0 0 0
G-e-f---r:I(
~ ~ ~
~
~:;
0
G.-.J.-·...[jJ--.--i
cO
00,0
c
00
a
BN
C(m)-B
NC(rn)
Cable
'-----n-rr
DC
BIAS
B
~ooo
[J
o
00000
o 0
DODD
0
o
o I = I
INTERFACE
BOX
TEST SIGNAL
Figure 1-14.
DC
Bias Level Accuracy Test Setup
Using an Interface
Box
MULTIMETER
I
cc
cc
o
ClOClOCJO
cc
000000
cc
.........
white
OCOc::J
OQCl
ccoc
OOco
Ou..
0000
000
red
~
black
2 BNCs(m)-3 alligator clips
Test Lead
___
* Direct connection no cable
white
red
HP
4285A
Dj
ClOCl
~ODD
ffl
B
MODO
[J
o 0
oOOCl
o 0
~ODD
0
CJ
o~
o
o I = I
Tee
BNC(m)(f)(f)
Adapter
Figure 1-15. DC Bias Level Accuracy Test Setup
Without an Interface
Box
EQUIPMENT:
Interface
Box
Multimeter
BNC(m)-to-BNC(m) Cable,
61
cm
BNC(f)
to
dual banana plug Adapter
1·27
HP
PN
04284-65007
HP 3458A
HP
PN
8120-1839
HP
PN