Specifications

PERFORMANCE
TESTS
2 m High Frequency Test
102. Move the cursor
to
the LEVEL field, and set the test signal level to 0.5
V.
103. Connect the Terminal Adapter to the
HP
4285A's
UNKNOWN
terminals using the
2 m Test Leads.
104. Store the OPEN CORRECTION data and the SHORT CORRECTION data, refer to
steps 48 to 54.
105. Press the 'DISPLAY FORMAT MENU' key to display the MEAS DISPLAY page.
106. Connect the 20
cm
Air Line to the Terminal Adapter's
OUTPUT
terminal.
107.
Connect the 0 S Termination to the 20 cm Air Line.
108. Move the cursor to the LEVEL field, and set the test signal level to 1
V.
109. Move the cursor to the
FREQ
field, and set the test frequency to
10
MHz.
110. Press the TRIGGER key, and confirm that the
HP
4285A's reading
is
within the
specified test
limits listed in Table 1-12.
111. Perform the test for
all settings listed in Table 1-12, by repeating steps 108 to
110 for each listed setting.
Table 1-12.
Impedance Measurement Accuracy Test Limits
(2
m High Frequency Test)
Test Test
Signal
Test Limits
Frequency Level
Cp
D
10 MHz 1
V
C.V.±0.208
pF
±0.0149
10
MHz 2 V C.V.±0.194 pF
±0.0139
30
MHz
1
V
C.V.±0.816
pF
±0.0583
30
MHz 2 V C.V.±0.789
pF
±0.0564
C.V.:Calibrated Value
at 1 MHz, using
an
HP
4284A
1-23