Specifications

PERFORMANCE
TESTS
6. Move the cursor to the
FREQ1
field, and press the 'MEAS OPEN' softkey to store
the open correction data.
7. Connect the
SHORT Termination to the
HP
4284A's UNKNOWN terminals.
8. Press the
'MEAS SHORT' soft key to store the short correction data.
9.
Press the MEAS
SETUP
MENU key to display the MEAS SETUP page.
10. Move the cursor
to
the
FREQ
field, and set the test frequency to 1.00000 MHz.
11. Move the cursor to the
TRIG
field, and press the
'MAN'
softkey to set the trigger
mode to manual.
12. Move the cursor to the
AVG
field, and press the 4 and ENTER keys
to
set the
averaging rate to 4.
13. Press the
DISPLAY FORMAT MENU key to display the MEAS DISPLAY page.
14. Connect the
10
pF
Standard to the
HP
4284A's UNKNOWN terminals.
15. Press the
TRIGGER
key.
16. Record the
Cp
reading into Calculation Sheet column
[b].
17.
Connect the Terminal Adapter
(HP
16085B) to the
HP
4284A's UNKNOWN
terminals.
18. Press the
MEAS
SETUP
MENU key and the 'CORRECTION' softkey
to
display the
CORRECTION
page.
19. Connect the
0 S Termination to the Terminal Adapter's OUTPUT terminal.
20. Move the cursor to the'
FREQ1
field, and
presS!
the 'MEAS OPEN' softkey to store
the
OPEN correction data.
21. Connect the
0 n Termination to the Terminal Adapter's OUTPUT terminals.
22. Press the
'MEAS SHORT' softkey to store the short correction data.
23. Press the
DISPLAY FORMAT MENU key to display the MEAS DISPLAY page.
24. Connect the
20 cm Air Line to the Terminal Adapter's OUTPUT terminal.
25. Connect the
0 S Termination to the 20 cm Air Line.
26. Press the
TRIGGER
key, and record the Cp reading into Calculation Sheet
column [c]. .
27. Calculate the Air Line calibration value for 1 MHz, according to the Calculation
Sheet.
1ยท16