Specifications

PERFORMANCE TESTS
7. Confirm that the Level Monitor reading
(Vm)
is
within the specified test limits
listed in Table 1-6.
8.
Set the test frequency and the test signal level according to Table 1-6, and con-
firm that the Level Monitor readings
are
within the specified test limits listed in
Table 1-6.
Table 1-6. Test Signal Level Accuracy Test Limits
Test
Test
Signal Test Limit
Frequency Level
100
kHz
1 V
0.919 V to 1.081 V
1 MHz
5 mV 3.580 mV to 6.52 mV
1 MHz 20 mV
17.32 mV to 22.68 mV
1 MHz
100 mV
90.6 mV to 109.4 mV
1
MHz
1 V 0.915 V to 1.085 V
1 MHz
2 V 1.831 V to 2.169 V
10 MHz
1 V
0.88 V to 1.12 V
30
MHz 1 V 0.80 V to 1.20 V
1-12