Specifications

PERFORMANCE TESTS
EQUIPMENT:
Interface
Box
Multimeter
BNC(m)-BNC(m)
Cable,
61
cm
BNC(f) to dual banana plug Adapter
PROCEDURE:
HP
PN
04284-65007
HP
3458A
HP
PN
8120-1839
HP
PN
1251-2277
1.
Set up the equipment
as
shown in Figure 1-2.
NOTE
If the Interface Box is not available, use the following cables and adapters
as
a substitute. Figure 1-3 shows the test setup without the interface box.
BNC(m)-BNC(m)
Cable, 30 cm
Tee, BNC(m)(f)(f) Adapter
2.
Set the multimeter
as
follows.
ACV
Auto
SETACV
SYNC
NPLC = 100
HP
PN
8120-1838 2
ea.
HP
PN
1250-0781
3. Perform a
SYSTEM
RESET
as
described on page 1-4.
4.
Press the MEAS
SETUP
MENU key and the 'CORRECTION' soft key to display the
CORRECTION
page.
5.
Move the cursor to the CABLE field, and press the
'0
m' softkey to set the cable
length to
0 m.
6.
Press the DISPLAY FORMAT MENU key to display the MEAS DISPLAY page.
7.
Move the cursor to the
FREQ
field and set the test frequency to 100 kHz.
8. Record the multimeter reading into Calculation Sheet column [a], and
calculate
the test limits according to the calculation sheet.
9.
Confirm that the Level Monitor reading (Vm)
is
within the specified test limits.
10. Perform the test for a" settings listed in Table 1-4 by repeating steps 7 to 9 for
each
listed setting.
1-7