Specifications

PERFORMANCE TESTS
1-7. TEST SIGNAL LEVEL MONITOR ACCURACY TEST
This test verifies that the
HP
4285A test signal level monitor accuracy
is
within
±[(4+O.2xfm)% of reading
+
Vm
x 0.002].
(fm:test frequency [MHz], Vm: test signal
level
setting) The accuracy
is
specified for
an
ambient temperature range of
23
0 C ±5 0
C.
The test consists of a Low Frequency Test and
a High Frequency Test.
7 -1. LOW FREQUENCY TEST
MULTIMETER
BNC(f)-dual
banana
plug
Adapter
o
l::lI:JOOOCJ
OClCOCJO
CO
ecce
o~t=1p---1
~ ~
~!;~~
OG
oc
cooo
000
BN
C(m)-B
N
C(m)
Cable
61
em
______
HP
4285A
D~
OOOOOOO:
o
ODD'
OM
0
o
00000
o 0
DODD
0
o 0 I = I
INTERFACE
BOX
DC
BIAS
TEST
SIGNAL
Figure 1-2. Test Signal
Level
Monitor Accuracy Test Setup
using
an
Interface Box
MULTIMETER
00
00
Cable
61
em
* Direct connection
no
cable
HP 4285A
I
DII~
oo~oorn
B 0
DOD
lhJ
o
10
01
DOD
o 0
~ODD
0
BNC(m)-BNC{m)
Cable
30
em
Adapter
.
o 0 I = I
BNC{m)-BNC{m)
Cable
30
em
Figure 1-3. Test Signal
Level
Monitor Accuracy Test Setup
without
an
Interface Box
NOTE
The
length of the
BNC
to
BNC
cables used in this test should
be
~
61
em.
1·6