1-6. TEST FREQUENCY ACCURACY TEST
This test verifies that the
HP
4285A test frequency accuracy
is
within ±0.01 %. The accuracy
is specified for
an
ambient temperature range of
23°C
±5
°C.
HP 4285A
ELECTRONIC
COUNTER
D~DDDD~~
8
~DDD
0
o
00
000
_ClClClOCICIO
o 0
0000
CI
CI
CI CI CI CI
CI
CI CI CI
CI
CICICICI
CI
CICICI
ClCICICIOO
o 0
000i
CI
CICIO e
e
e
0
CI
I = I
.........
L...-J
L---I
BNC(m)-BNC(m)
Cable.
61
em
Figure 1-1. Test Frequency Accuracy Test Setup
EQUIPMENT:
Electronic Counter
BNC(m)-to-BNC(m)
Cable,
61
cm
PROCEDURE:
HP
5334B
HP
PN
8120-1839
1.
Set up the equipment
as
shown in Figure 1-1.
2. Perform a
SYSTEM
RESET
as
described on page 1-4.
3. Move the cursor to the
FREQ
field.
L--J
4. Set the test frequency according to Table 1-3, and confirm that the counter read-
ings are within the test limits given in the table.
Table 1-3. Test Frequency Test limits
Test Frequency Test Limits
100 kHz 99.990 kHz to
100.010 kHz
149 kHz 148.985 kHz to
149.015 kHz
1 MHz
0.99990
MHz to 1.00010 MHz
10 MHz 9.9990
MHz
to
10.0010 MHz
16
MHz
15.9984
MHz to 16.0016
MHz
17
MHz 16.9983
MHz
to
17.0017 MHz
30
MHz 29.9970
MHz to 30.0030
MHz
1-5