Specifications
1-4.
INITIAL
OPERATING PROCEDURE
Before starting the performance tests, the following procedure must be performed.
[Procedure]
1. Turn the
HP
4285A ON.
NOTE
Steps 2 through 4 must
be
performed only when the
HP
4285A
is
equipped with Option 002 Accessory Control Interface.
2. Press the CATALOG SYSTEM key and
the
'SYSTEM
CON
FIG' softkey to
display the
SYSTEM
CON
FIG
page.
3. Move the cursor to the
CON
FIG
field, using the
CURSOR
arrow keys.
4. Press the 'OFF' softkey to disable the Accessory Control
Interface.
5. Warm-up the
HP
4285A for at least 30 minutes to allow its internal circuits to
stabilize.
1-5. SYSTEM RESET
By using SYSTEM RESET function the
HP
4285A can
be
set up easily for the performance
tests.
SYSTEM RESET can
be
performed using the following procedure.
[Procedure]
1 . Press the
MEAS
SETUP
MENU key to display the MEAS SET
UP
page.
2. Move the cursor to the
SYS
MENU field, using the
CURSOR
arrow keys.
3. Press the 'more 1/2' softkey to display the
'SYSTEM RESET' softkey.
4. Press the
'SYSTEM RESET' softkey.
5. Press the
'YES' softkey to perform a
SYSTEM
RESET.
1-4