Specifications
○
C
Copyright 2012 Agilent Technologies
Change 15
Revised Minimum, Maximum & Measurement Uncertainty in DC BIAS LEVEL ACCURACY TEST (OPTION
001 ONLY).
Bias Voltage
Minimum
Measured
Value
Maximum
Measurement
Uncertainty
0.0 V
-0.0010 V
0.0010 V
0.001 mV
0.1 V
0.0989 V
0.1011 V
0.011 mV
2 V
1.9970 V
2.0030 V
0.022 mV
6 V
5.9920 V
6.0080 V
0.062 mV
14 V
13.981 V
14.019 V
0.198 mV
30 V
29.960 V
30.040 V
0.390 mV
40 V
39.950 V
40.050 V
0.510 mV
-0.1 V
-0.1011 V
-0.0989 V
0.011 mV
-2 V
-2.0030 V
-1.9970 V
0.022 mV
-6 V
-6.0080 V
-5.9920 V
0.062 mV
-14 V
-14.019 V
-13.981 V
0.198 mV
-30 V
-30.040 V
-29.960 V
0.390 mV
-40 V
-40.050 V
-39.950 V
0.510 mV
Change 16
Revised Minimum, Maximum & Measurement Uncertainty in VOLTAGE RATIO MONITOR ACCURACY
TEST (OPTION 002 ONLY).
C.V.(Step Attenuator 20dB Cal. Value at 1 MHz): [dB]
Minimum
Measured Value
Maximum
Measurement
Uncertainty
C.V. - 0.087 dB
C.V. + 0.086 dB
0.022 dB