Specifications

C
Copyright 2012 Agilent Technologies
Change 13
Revised Minimum, Maximum & Measurement Uncertainty in Impedance Measurement Accuracy Test
(1000pF STANDARD (2 m)).
1000 pF STANDARD (2 m) C.V.: (Cal.Value at 1MHz)
Test Frequency: 1MHz
Upper Value: Capacitance (Cp)
Lower Value: Dissipation Factor (D)
Test Signal
Level
Minimum
Measured
Value
Maximum
Measurement
Uncertainty
1 V
C.V. - 2.19 pF
C.V. + 2.19 pF
0.5000 pF
-0.0022
0.0022
0.0002
2 V
C.V. - 2.55 pF
C.V. + 2.55 pF
0.5000 pF
-0.0025
0.0025
0.0002
Change 14
Revised Minimum, Maximum & Measurement Uncertainty in Impedance Measurement Accuracy Test (20
cm AIR LINE (2 m)).
20 cm AIRLINE (2 m) C.V.: (Cal.Value at 1MHz)
Test Frequency: 1MHz
Upper Value: Capacitance (Cp)
Lower Value: Dissipation Factor (D)
Test Frequency
Test Signal Level
Minimum
Measured Value
Maximum
Measurement
Uncertainty
10 MHz
1 V
C.V. - 0.2022 pF
C.V. + 0.2022 pF
0.0230 pF
-0.0150
0.0150
0.0016
10 MHz
2 V
C.V. - 0.1889 pF
C.V. + 0.1889 pF
0.0230 pF
-0.0140
0.0140
0.0016
30 MHz
1 V
C.V. - 0.7903 pF
C.V. + 0.7903 pF
0.1500 pF
-0.0585
0.0585
0.0105
30 MHz
2 V
C.V. - 0.7640 pF
C.V. + 0.7640 pF
0.1500 pF
-0.0565
0.0565
0.0105