Specifications
○
C
Copyright 2012 Agilent Technologies
Change 9
Revised Minimum, Maximum & Measurement Uncertainty in Impedance Measurement Accuracy Test
(100pF STANDARD (1 m)).
100 pF STANDARD (1 m) C.V.: (Cal.Value at 1MHz)
Test Frequency: 1MHz
Upper Value: Capacitance (Cp)
Lower Value: Dissipation Factor (D)
Test Signal Level
Minimum
Measured
Value
Maximum
Measurement
Uncertainty
1 V
C.V. - 0.220 pF
C.V. + 0.220 pF
0.0250 pF
-0.0022
0.0022
0.0002
2 V
C.V. - 0.285 pF
C.V. + 0.285 pF
0.0250 pF
-0.0029
0.0029
0.0002
Change 10
Revised Minimum, Maximum & Measurement Uncertainty in Impedance Measurement Accuracy Test
(1000pF STANDARD (1 m)).
1000 pF STANDARD (1 m) C.V.: (Cal.Value at 1MHz)
Test Frequency: 1MHz
Upper Value: Capacitance (Cp)
Lower Value: Dissipation Factor (D)
Test Signal Level
Minimum
Measured
Value
Maximum
Measurement
Uncertainty
1 V
C.V. - 2.19 pF
C.V. + 2.19 pF
0.5000 pF
-0.0022
0.0022
0.0002
2 V
C.V. - 2.55 pF
C.V. + 2.55 pF
0.5000 pF
-0.0025
0.0025
0.0002